Background

Stent dislodgement is a life-threatening complication that can result in coronary artery embolization, stent thrombosis, acute myocardial infarction, and even death. Severely angulated, heavily calcified, and previously stented coronary arteries are associated risk factors. With the development of different lesion preparation techniques and the drug eluting stent era, the reported incidence of stent dislodgement has decreased to < 1% in the last few years.

Case presentation

We report a case of a 64-year-old Moroccan man complicated during percutaneous intervention in the left main artery by the loss of two stents. This complication was successfully managed by passing the stent’s balloon into the stent and then fully expanding it. In our case, the device’s characteristics were involved and could play a role in such complications, but it is still not well understood.

Conclusions

The main treatment option is stent retrieval with different available techniques. If retrieval of the stent is impossible, crushing it against the blood vessel wall could be considered.

Peer Review reports

Background

Stent dislodgement is a challenging and serious complication during percutaneous intervention (PCI). It is a life-threatening complication that can result in coronary artery embolization, stent thrombosis, acute myocardial infarction, and even death. Moreover, if a lost stent moves outside of the coronary artery, it may cause cerebral stroke, and peripheral artery occlusion [1].

Several lesion characteristics are related to stent dislodgement, such as heavily calcified stenosis, severe tortuosity, long diffuse lesions, and previously implanted stents. Other stent-related factors may be involved such as stent designs, stent strut thickness, and metal platforms [2]. With the development of different lesion preparation techniques and the drug eluting stent era, the reported incidence of stent dislodgement has decreased to < 1% in the last few years [3]. But unfortunately, it still occurs; therefore, interventional cardiologists must be familiar with common stent retrieval techniques.

We report a complicated case of the loss of two stent that occurred during PCI in the left main coronary artery, and also demonstrate a safe and successful approach to deal with this issue.

Case presentation

A 64-year-old Moroccan man with a medical history of an acute coronary syndrome 2 months previous leading to angioplasty of the mid and distal left anterior descending artery (LAD) with two drug-eluting stents. The previous coronarography showed a distal left main stenosis. After medical surgical discussion and considering the low syntax score with no diabetes, the patient was admitted in the cathlab for a scheduled distal left main angioplasty. The patient had persistent Canadian class III angina on optimal medical therapy. He was hemodynamically stable on physical examination with a blood pressure at 128/72 mmHg, a pulse at 90 beats per minute, and saturation at 96% on room air. The physical exam was normal. The lungs were clear to auscultation, there was no leg edema, and peripheral pulses were palpable. Electrocardiogram showed a regular sinus rhythm with anterior Q waves. The troponin level and usual biological assessment was normal. Echocardiography found features of ischemic heart disease with anterior and apical hypokinetic wall motion and a preserved left ventricular ejection fraction at 50%, normal left ventricle feeling pressure, and no mitral regurgitation or pulmonary hypertension.

The patient underwent angioplasty via the right femoral artery through 7-French femoral sheath. There was a 70-90% stenosis of the distal left main (LM) coronary artery including proximal LAD and circumflex (LCx; MEDINA 1.1.1). Using a 7 French extra back up guiding catheter well engaged in the left main, both LAD and LCx were wired with workhorse wires Sion blue and BMW universal II respectively. PCI was attempted from the left main to the proximal stenosis of LAD initially with a 4 × 26 mm Orsiro Biotronik Drug-Eluting Stent (DES). The stent was dislodged just after the exit of the guiding catheter with guidewire still maintained in the LAD (Fig. 1).

Fig. 1
figure 1

Stent loss in the proximal LAD (red arrow)

Since the guidewire position was maintained through the lost stent, an attempt to advance the stent’s balloon through the lost stent was performed. The balloon could pass, and about 70% of the dislodged stent was deployed (Fig. 2A). Other small balloons were used to expand the distal part of the stent. After post-dilatation with a final balloon 4.00 × 15 mm, the whole stent was successfully deployed and covered 4 mm in the distal left main to proximal LAD (Fig. 2B).

Fig. 2
figure 2

A Successful deployment of proximal and middle part of the stent of proximal LAD (red arrow), B Post-dilatation of the distal part of the stent to ensure a full expansion (blue arrow)

The guide wires were switched to ensure that they were not twisted. The LCx was predilated with a 3.00 × 15 mm compliant balloon. Then another Orsiro Biotronik 4.0 × 22 mm was chosen to treat the left main–circumflex stenosis. But similarly to the first stent, at the exit of the guide catheter, it was partially dislodged in the left main–circumflex (Fig. 3A). The attempt to retrieve the stent was unsuccessful, and then the stent was deployed in the left main after advancing the stent’s balloon. But unfortunately, the stent failed to completely cross and cover the circumflex lesion. The proximal circumflex was then treated with a 4.00 × 15 mm Promus Elite Boston stent without any issue. A 3.00 × 15 mm compliant balloon was used to open struts in the left main–circumflex, then full stent expansion was obtained by a final proximal optimization technique (POT). The final angiography showed a fairly good result with no residual stenosis (Fig. 3B). The patient was regularly followed up with complete resolution of chest pain. A systematic angiogram was performed 7 months after, which found a satisfying result with no restenosis (Fig. 4). Stent boost and intravascular ultrasound (IVUS) were not available.

Fig. 3
figure 3

A Partial stent loss in the left main–ostial circumflex, B Final result after stenting the ostial circumflex

Fig. 4
figure 4

Angiogram at 7 months showing a good result with no significant restenosis

Discussion

The incidence of stent loss has significantly declined in recent years despite the high number of complex PCI procedures performed. The incidence of stent loss in studies published before 2000, between 2000 and 2005, and after 2005 was 5%, 1%, and 0.3%, respectively, as shown in Table 1 [4,5,6,7]. This may reflect the continuous improvement of stent technology, including pre-mounted stents, better cross-sectional profiles, and improved stent delivery systems, as well as the increasing experience of operators.

Table 1 Decrease in incidence of stent loss in different studies between 1998 and 2020

Several risk factors for stent loss have been proposed by previous studies. As reported by Laarman et al., direct stenting may be associated with a higher risk of stent loss compared with a strategy of pre-dilation, presumably due to the increased resistance to stent advancement through the lesion [8]. The other most common causes of stent dislodgement are attempting to deliver a stent through a previously deployed stent. Stenting from distal to proximal may prevent such a complication. Sometimes, stent deformation can occur during attempts to cross the lesion, and then the stent is stripped off while attempting to withdraw it into the guide catheter [9, 10]. Tortuous vessels and heavily calcified lesions are other factors that could contribute to stent loss [11, 12]. However, the development of preparation lesion techniques (such as predilation, orbital or rotational atherectomy, and shockwave) has allowed for better stent delivery, and therefore lowered the risk of stent loss. In fact, adequate and careful preparation of the stenting site, such as the reduction of calcium burden, intuitively reduces the risk of stent decrimping, dislodgement, or embolization. In our patient, the lesion was not heavily calcified or very tight; that is why a direct stenting approach was chosen and justified. Moreover, the guide catheter was well engaged in the left main with no guide wire twist, and the stent loss happened just after the exit from the guide catheter and not in the attempt to cross the lesion. In this case, a lack of lesion preparation may play a role, but since the stent dislodgement happened just at the exit of the guide catheter and not in the attempt to cross the lesion, stent-related factors may be involved in the stent dislodgement in our case.

Latest reports suggest that other factors related to the stent may increase the risk of stent dislodgement. Kyoung Woo Seo and his team compared dislodgement forces in five different types of DES: Firehawk, Xience Sierra, Orsiro, Resolute Onyx, and Synergy. They found that the peak dislodgement force was significantly lower in the Orsiro than in all the other stents. Moreover, during the pullback of the stents, other DESs except the Orsiro were not removed from the delivery system despite the stents being broken. The Orsiro stent was easily removed if a strut was dislodged from the balloon [13]. Rigatelli et al. reported on the relationship between stent strut thickness and dislodgement. They divided the stents into thick (> 81 µm strut thickness) and ultrathin (≤ 81 µm strut thickness) strut stent groups. Stent dislodgement is more common in ultrathin than in thick strut stents (0.28% versus 0.78%, P < 0.001). The ultrathin strut group included Resolute Onyx, Orsiro, Xience, and Coroflex. Although they did not report the incidence of each stent, approximately half of the stents were Orsiro stents, which had the thinnest strut, and this group displayed a higher rate of stent dislodgement [7, 14]. In the same report, the authors explain that, even if the rate of dislodgement is higher with ultrathin stents, they could be easily parked or crushed with a lower risk of stent thrombosis. This may suggest that the use of a thin strut DES in our patient could have played a role in stent dislodgement. Moreover, the use of another thick strut stent without any problem in our case supports these findings.

Few systematic studies have been conducted on the incidence and outcomes of stent dislodgement in patients treated with DES over the last 20 years. In 2007 Yucel Colkesen et al. observed a 0.29% incidence of stent embolization in a series of 4797 patients that resulted in bypass surgery in most cases [15]. Alomar et al., in a meta-analysis across the 1991–2012 period, reported stent loss in 1.3% of 71,655 PCI, resulting in a 19% complication rate including coronary artery bypass graft surgery, 18% for myocardial infarction, a 19% death rate, 6% for bleeding requiring transfusion, 3% for vascular access complications, 0.6% rate of cerebrovascular accident [3]. These complications appear to be less frequent in a recent report from 2016 where the incidence of stent loss was only 0.56% out of 25,692 PCI. Concerning complications, there was no acute death or need for urgent coronary artery bypass graft (CABG), and they were all managed, including through vessel dissection, coronary perforation, stent thrombosis, and acute myocardial infarction [7].

Appropriate management of the dislodged stent is important to prevent those complications. Several management options are available for dealing with this critical situation and include: (1) the small-balloon technique—only feasible if the guidewire position is maintained through the lost stent—which is a simple technique in which a small balloon is advanced through the lost stent, inflated distally, and withdrawn, displacing the lost stent into the guiding catheter [6]; (2) retrieving the stent with snare; (3) leaving the stent in the coronary vessel and crushing it with another stent [16, 17]; and (4) parking the stent in a peripheral vessel in the case of extra-coronary migration [18, 19]. Actually, most of the cases are successfully managed via a percutaneous approach, while surgical intervention is required mainly if the stent embolizes and causes coronary occlusion leading to hemodynamic instability. However, peripheral stent loss rarely causes complications, and in many cases, it cannot be localized. In our case, since we could maintain the guidewire position through the lost stent, we could repass the stent balloon and deploy and fully expand the stent.

Conclusions

Stent dislodgement is a rare but serious complication. Severely angulated, heavily calcified, and previously stented coronary arteries are associated risk factors. It is probable that the device’s characteristics could play a role in such complications, but it is still not well understood. The main treatment option is stent retrieval with different available techniques. If retrieval of the stent is impossible, crushing it against the blood vessel wall could be considered.

Availability of data and materials

The published information is available from the corresponding author on reasonable request.

References

  1. Bolte J, Neumann U, Pfafferott C, Vogt A, Engel HJ, Mehmel HC, et al. Incidence, management, and outcome of stent loss during intracoronary stenting. Am J Cardiol. 2001;88(5):565–7.

    Article  CAS  PubMed  Google Scholar 

  2. Iturbe JM, Abdul-rahman R, Papayannis A, Mahmood A, Rangan BV, Banerjee S, et al. Frequency, treatment, and consequences of device loss and entrapment in contemporary percutaneous coronary interventions. J Invasive Cardiol. 2012;24(5).

  3. Alomar ME, Michael TT, Patel VG, Altomare CG, Rangan BV. Stent loss and retrieval during percutaneous coronary interventions: a systematic review and meta-analysis. J Invasive Cardiol. 2013;25(12).

  4. Brilakis ES, Best PJ, Elesber AA, Barsness GW, Lennon RJ, Holmes DR Jr, et al. Incidence, retrieval methods, and outcomes of stent loss during percutaneous coronary intervention: a large single-center experience. Catheter Cardiovasc Interv. 2005;66(3):333–40.

    Article  PubMed  Google Scholar 

  5. Cantor WJ, Lazzam C, Cohen EA, Bowman KA, Dolman S, Mackie K, et al. Failed coronary stent deployment. Am Heart J. 1998;136(6):1088–95.

    Article  CAS  PubMed  Google Scholar 

  6. Eggebrecht H, Haude M, Von Birgelen C, Oldenburg O, Baumgart D, Herrmann J, et al. Nonsurgical retrieval of embolized coronary stents. Catheter Cardiovasc Interv. 2000;51(4):432–40.

    Article  CAS  PubMed  Google Scholar 

  7. Rigatelli G, Zuin M, Vassilev D, Dell’Avvocata F, Giordan M, Conte L, et al. Risk of dislodgement of ultrathin drug eluting stents versus thick drug eluting stents. Am J Cardiol. 2020;125(11):1619–23.

    Article  PubMed  Google Scholar 

  8. Laarman G, Muthusamy TS, Swart H, Westendorp I, Kiemeneij F, Slagboom T, et al. Direct coronary stent implantation: safety, feasibility, and predictors of success of the strategy of direct coronary stent implantation. Catheter Cardiovasc Interv. 2001;52(4):443–8.

    Article  CAS  PubMed  Google Scholar 

  9. IJsselmuiden AJJ, Tangelder GJ, Cotton JM, Vaijifdar B, Kiemeneij F, Slagboom T, et al. Direct coronary stenting compared with stenting after predilatation is feasible, safe, and more cost-effective in selected patients: evidence to date indicating similar late outcomes. Int J Cardiovasc Intervent. 2003;5(3):143–50.

    Article  CAS  PubMed  Google Scholar 

  10. Holmes DR. Stent complications. J Invasive Cardiol. 1998;10(7):385–95.

    PubMed  Google Scholar 

  11. Kostantinis S, Karacsonyi J, Simsek B, Brilakis ES. Complications of stent loss during treatment of a heavily calcified and tortuous chronic total occlusion. Cardiovasc Revasc Med. 2022;40:293–7.

    Article  Google Scholar 

  12. Lohavanichbutr K, Webb JG, Carere RG, Solankhi N, Jarochowski M, Yulia D, et al. Mechanisms, management, and outcome of failure of delivery of coronary stents. Am J Cardiol. 1999;83(5):779–81.

    Article  CAS  PubMed  Google Scholar 

  13. Seo KW, Yang HM, Lim HS, Yoon MH. Stent dislodgement force of drug-eluting coronary stents: a bench test. Cardiovasc Diagn Therapy. 2022;12(3):370.

    Article  Google Scholar 

  14. Rigatelli G, Zuin M, Gianese F, Adami D, dell’Avvocata F, Barison S, et al. Ultrathin biodegradable-polymer orsiro drug-eluting stent performance in real practice challenging settings. Cardiovasc Revasc Med. 2021;30:12–7.

    Article  PubMed  Google Scholar 

  15. Colkesen AY, Baltali M, Acil T, Tekin G, Tekin A, Erol T, et al. Coronary and systemic stent embolization during percutaneous coronary interventions a single center experience. Int Heart J. 2007;48(2):129–36.

    Article  PubMed  Google Scholar 

  16. Wongpraparut N, Yalamachili V, Leesar MA. Novel implication of combined stent crushing and intravascular ultrasound for dislodged stents. J Invasive Cardiol. 2004;16(8):445–6.

    PubMed  Google Scholar 

  17. Kirk MM, Herzog WR. Deployment of a previously embolized, unexpanded, and disarticulated Palmaz-Schatz stent. Cathet Cardiovasc Diagn. 1997;42(3):331–4.

    Article  CAS  PubMed  Google Scholar 

  18. Egbuche O, Mezue KN, Nwokike SI, Abe T, Olanipekun T, Onuorah I, et al. Left main stenting with stent dislodgement and entrapment in the common femoral artery: a successful transcatheter stent retrieval. Am J Cardiovasc Dis. 2021;11(3):421.

    PubMed  PubMed Central  Google Scholar 

  19. Meisel SR, DiLeo J, Rajakaruna M, Pace B, Frankel R, Shani J. A technique to retrieve stents dislodged in the coronary artery followed by fixation in the iliac artery by means of balloon angioplasty and peripheral stent deployment. Catheter Cardiovasc Interv. 2000;49(1):77–81.

    Article  CAS  PubMed  Google Scholar 

Download references

Acknowledgements

No acknowledgements for this article.

Funding

This work was not supported by any funding.

Author information

Authors and Affiliations

Authors

Contributions

All authors were involved in the diagnosis, management, and care of the patient. All authors contributed equally to writing the manuscript.

Corresponding author

Correspondence to Salim Arous.

Ethics declarations

Ethics approval and consent to participate

Not applicable.

Consent for publication

Written informed consent was obtained from the patient for publication of this case report and any accompanying images. A copy of the written consent is available for review by the Editor-in-Chief of this journal.

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Arous, S., Zahidi, H., El Ghali Benouna, M. et al. Two stents’ dislodgement in the left main coronary artery: a case report. J Med Case Reports 18, 158 (2024). https://0-doi-org.brum.beds.ac.uk/10.1186/s13256-024-04491-6

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://0-doi-org.brum.beds.ac.uk/10.1186/s13256-024-04491-6

Keywords